FP-30 / FP-40 Formaldehyde Gas Detector

Specific to Formaldehyde (HCHO) in air.

 

Aktuelles Datenblatt

FP-30-40

FP-30 / FP-40

These highly sensitive portable gas detectors use photoelectric photometry method with colorimetric detection tabs for a safe and fresh measurement at all times. Moreover, the sample drawing rate with flow control is controlled by a microprocessor.
Both the FP-30 and FP-40 have a direct readout of measurement value on a large LCD display and a wide range of self-diagnosis features for all safety-relevant functions.
For best monitoring and traceability FP-40 comes with data storage for up to 99 readings and PC-interface for downloading and analyzing prior measurements.

 

Features

  • Specific to Formaldehyde (HCHO) in air
  • Stores 99 readings, direct digital readout
  • No warm up time
  • Small and light, only 500g
  • Built-in sample draw pump
  • Self-diagnosis

 

Model

FP-30

FP-40

Detection gas

Formaldehyde (HCHO) in air

Detection range

0 ~ 0.4 ppm
(0.005 ppm/digit)

0 ~ 1.0 ppm
(0.01 ppm/digit)

0 ~ 0.6 ppm
(0.005 ppm/digit)

Detection time (Intervall)

1800 seconds (30 minutes)

900 seconds (15 minutes)

180 seconds (3 minutes)

Detection method

Photoelectric photometry method
(colorimetric tablet method, accumulating measurement)

Sampling method

Sample drawing with built-in pump

Data Logger

Detection point: up to 99 points
(automatic recording at the completion of measurement)

Display

Backlighted digital LCD display

Self-diagnosis functions

Failure of light source and light receiver, low battery voltage, pump failure, system trouble

Alarm indication

LCD display, buzzer sound

Operating conditions

-10°C ~ +40°C, below 90% RH (non-condensing)

+5°C ~ +35°C, below 90% RH
(non-condensing)

Continuous operation

Approx. 12 hours (without alarm and backlight, with alkaline batteries at 20°C)

Power source

AA size Alkaline batteries (4x 1,5V)

Dimensions and weight

Approx. 85 x 190 x 40 mm, 500 g

Accessories

Standard:     20 detection tabs, carrying case, AA size alkaline batteries, operation manual

Optional:    Filter, data logger software and PC connection cable (FP-40), tripod, dust filter

Specifications subject to change without notice.